Channel volt |
Measurement range SENS |
±0.1 / 0.2 / 0.5 / 1 / 2 / 5 / 10 / 20 / 30 / 50 / 100 V |
Accuracy at ambient temperature 25 °C |
±0.15 % (unipolar measurement ranges), ±0.10 % (bipolar measurement ranges) |
Drift for ambient temperature -40 ... 85°C |
±40 ppm/K |
Drift for ambient temperature 85 … 105 °C |
±80 ppm/K |
Drift for ambient temperature 105 … 125 °C |
±250 ppm/K |
Input HV |
* Safety instructions |
|
Channel current |
Measurement range |
0 … 20 mA, ±20 mA |
Accuracy |
±0.50 % |
Internal shunt resistor |
50 Ω |
General channel properties |
Special functions |
Offset adjust, during measurement, multiple groups |
AD converter |
16 bit / SAR (successive approximation register) |
Oversampling |
2 kHz |
Channel sampling rates |
1 / 2 / 5 / 10 / 20 / 50 / 100 / 200 / 500 / 1000 / 2000 Hz |
Aggregate sample rate |
16 kHz |
Hardware filter (switchable) |
150 Hz (M-SENS 8 / M-SENS 8 DSP), Accuracy 10 % |
Software filter types |
Butterworth, Elliptic (8-pole), Bessel |
Channel impedance |
10 MΩ |
Software filter (DSP selectable) |
6 / 7.5 / 9.96 / 15 / 30 / 39.96 min, 16.67 / 25 / 50 / 66.7 / 100 / 125 Hz (M-SENS 8 DSP), Accuracy 0.05 %, 1 / 1.25 / 1.67 / 2.5 / 5.0 / 6.67 / 10 / 12.5 Hz |
Channel LED |
Yes, Yes, Channel LED is flashing during configuration |
TEDS |
Class 2 |
Channel CNT |
Number of inputs |
|
Excitation |
Sensor excitation ranges |
Bipolar ±2.5 / ±5 / ±7.5 / ±8 / ±10 / ±12.5 / ±15 V |
Sensorspeisung |
|
Accuracy excitation at ambient temperature 25 °C |
±0.30 % |
Accuracy excitation at ambient temperature 85 °C |
±0.50 % |
Accuracy excitation at ambient temperature 120 °C |
±0.70 % |
Sensor excitation current |
45 mA (for V output ±7.5 / ±15.0 V), 40 mA (for V output ±5.0 / ±12.5 V), 30 mA (for V output ±2.5 / ±10.0 V |
Galvanic isolation |
Input ↔ module power supply |
±100 V (indefinitely), ±200 V (short-time, t < 2 ms) |
Input ↔ CAN |
±100 V (indefinitely), ±200 V (short-time, t < 2 ms) |
Input ↔ enclosure |
±100 V (indefinitely), ±200 V (short-time, t < 2 ms) |
Input ↔ input |
±100 V (indefinitely), ±200 V (short-time, t < 2 ms) |
Input ↔ excitation |
±100 V (indefinitely), ±200 V (short-time, t < 2 ms) |
Device |
Inputs |
8 |
Precisions |
|
Maximum input protection voltage (channel) |
±100 V (indefinitely), ±200 V (short-time, t < 2 ms) |
Voltage supply |
9 … 36 VDC |
Supply voltage thresholds |
On 9 ±0.3 VDC / Off 6 ±0.3 VDC |
Power consumption, typical |
3.5 W (all excitations off) |
Working temperature range |
-40 … 125 °C (-40 … 257 °F) |
Storage temperature range |
-55 … 150 °C (-67 … 302 °F) |
IP-Code |
IP 67 (ISO 20653 - 2013) |
Relative humidity |
5 ... 95 % |
Dimensions |
W204 mm x H41 mm x D55 mm (8.03 in x 1.61 in x 2.17 in) |
Weight |
695 g (1.53 lb) |
Configuration interface |
CAN high speed |
Data transfer rate |
Software selectable up to 1 MBit/s (ISO11898-2) |
Housing material |
Aluminum, gold anodized |
Input sockets |
ODU series F, size 1 (5-pin) S11F1C-T05MJG0-2500, Lemo EGG 1B 307 (7-Pin) |
Status LED |
Yes |
Derating (decrease of total output power) |
-1.25 % /K for ambient temperature > 85 °C |
Schutz gegen aggressive Flüssigkeiten |
|
Electrical data |
Status LED |
|
Accessories |
System cable |
620-561 M-PWR term cable, banana, 620-567 M-CAN/PWR term cable SubD/S, banana, 620-560 M-CAN cable, 620-502 M-CAN cable SUBD/S Term., M-CAN-ABS |
Input cable |
670-810 SENS LEMO 1B 7p cable open, 620-674 SENS LEMO 1B6p I-measure cable open, 670-811 SENS LEMO 1B 7p TEDS cable open, 670-807 SENS LEMO 1B 6p cable open, 600-866 SENS LEMO 1B7p cable BNC/P, 600-731 SENS LEMO 1B 7p cable banana |